Date: |
2014-06-25 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Madeleine and I calibrated the 6400 in the lab
Zero: CO2 1.45 to 0.0
H2O -0.21 to -0.001
Span CO2 with 421.22 CO2 gas
--did not need to span--left at 421.0
H2O span (dew point set at 15.26C)
14.63 to 15.26C |
Entered by: |
Patty Oikawa |
|
|