Date: |
2014-03-07 |
From: |
Twitchell Rice - Gas Analyzer CH4 |
To: |
Twitchell Rice - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: Twitchell
Last calibration: 2013-08-21
RSSI=63%, Optical RH=17%
Ambient P=102.1kPa, Tair=18.5C
Zero Air at 1600psi, 2nd CH4 4.995 at 1100
CH4 Zero Offset: 0.01ppm
CH4 Span Offset: 0.004ppm
Notes: Calibrated by Cove and Sara. Firmware upgraded to v1.0.21 after calibration. Windy day.
|