Equipment and Calibration

View History

Equipment Entry

Instrument:Moisture Point
Manufacture:Environmental Sensors Inc.
Model No:PRB-K Type #8
Serial No:000000A6D8356
Tag No:
Acquisition Date:0000-00-00
Project:NSF-Methane
Location on 2010-09-03:Disposed,
Notes:

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
198 2010-09-03 Factory Disposed Disposed Report from vendor not repairable:
s/n 00000A6D8356
Condition on arrival: cracks in the epoxy of top segment (segment 1), top part of the probe head plastic is missing, missing a pin (diode 1 control pin). Ctrl Ground pin open circuit, as are all diode control lines. No valid diode waveforms from any of the 5 diode boards. Probe scans fail (all segments return error readings).
174 2010-07-07 Lab none Repair Factory Sent to ESI - Environmental Sensors Inc (www.esica.com)for evaluation and possible repair.
149 2010-06-22 Sherman Island Removed Lab none This probe was removed from the Sherman Island cow pasture. It was outside the cow fence giving bad readings. Second pin from the top is broken.