Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
200 | 2010-09-03 | Factory | Disposed | Disposed | Report from vendor not repairable:
s/n 000000573FAA Condition on arrival: Visual inspection shows only a missing pin (ID pin, which would not affect probe readings) and a very slight separation between epoxy and the plastic probe head. Ctrl ground and RF ground resistance checks fail, diode 2 resistance passes, but both diode 0 and diode 1 lines indicate an open circuit to one diode in each pair. Only 3 of the expected 5 diode waveforms observed in, probe scans fail (only 2 of 4 segments return values). |
176 | 2010-07-07 | Lab none | Repair | Factory | Sent to ESI - Environmental Sensors Inc (www.esica.com)for evaluation and possible repair. |
147 | 2010-06-24 | Vaira Moisture Point Solar | Removed | Lab none | Removed from Vaira between the solar panels and the diffuse radiation tripod. The top pin in the connector is broken. |
4238 | 2001-11-10 | Lab none | Installed | Vaira Moisture Point Solar | Installed at Vaira near the solar panels by Dennis and Liukang |