View history entry 1049 for Licor LI-7700 sn: TG1-0211

Date: 2013-08-21
From: Twitchell Rice - Gas Analyzer CH4
To: Twitchell Rice - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Twitchell Last calibration: 2013-02-21 RSSI=64%, Optical RH=13% Ambient P=101.2kPa, Tair=18.3C Zero Air at 1750psi, CH4 in air 4.995 at 1300 CH4 Zero Offset: 0.01ppm CH4 Span Offset: -0.025ppm Notes: It was windy but seemed to have no effect. The concentration of the secondary ch4 standard has been reevaluated since last cal. Last time 4.982, now 4.995