Equipment and Calibration History

View Equipment |

View history entry 1005 for Los Gatos Research FMA RMT-200 sn: 05-0010

Date: 2013-06-17
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2012-03-18 Cell Pressure=156.27Torr, Cell Temperature=24.76C, Ring Time=12.762usec Primary CH4 in Air at >1900 psi CH4 Span=1.89919 ppm CH4 = 1.895 -> 1.899 Offset: -4ppb Notes: Done in the lab with primary NOAA GMD standard. Cross cal with secondary CH4 tank = 4.995 ppm (n = 164). Placed new label on secondary CH4 tank. Last cross cal of 4.982 ppm on June 2, 2010.
Entered by: