Equipment and Calibration

Equipment history at Mayberry

| ALL | | Anemometer || Barometer || Battery Bank || Camera || CH4 Probe || CO2 Probe || CO2 Transmitter || Conductivity Sensor || Data Logger || DO Sensor || Fluorometer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || LED sensor || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Water Level |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
Precipitation
2015-06-25 40415-1006 TR-5251 Installed Precipitation Installed on a post along the boardwalk to the tower at Mayberry.
2018-10-25 40415-1006 TR-5251 Tested Precipitation Repaired chewed cable and tested with 475ml water. Got 100 tips. Did not adjust set screws.
2019-10-22 40415-1006 TR-5251 Calibration Precipitation Initially, the ground wire at the data logger was broken. After cleaning the tipping bucket we connected it to a CR23x and dripped 475ml of water through the bucket which should result in 100tips. We got 109 tips. We turned its set screw to full rotations CCW. Retest on 10-31 got 102 tips - good enough.
2021-06-22 40415-1006 TR-5251 Removed Precipitation Removed from MB and brought back to the lab for repair. It had lots of tips even though there hasn't been any rain, probably because the spliced area of the cable was underwater.
2021-07-06 40415-1006 TR-5251 Installed Precipitation Installed at MB again after replacing the cable in the lab.
2021-10-21 40415-1006 TR-5251 Tested Precipitation We dripped 240ml of water through the bucket using a peristaltic pump powered with a lithium battery. This should result in 50 tips in about 20 minutes. We got 54 tips and adjusted its set screws 1.5 turns CCW (1/2 turn of both screws is a 2-3% adjustment).
2021-11-04 40415-1006 TR-5251 Adjusted Precipitation Tested tipping bucket - got 53/50 tips. Adjusted screws 1/2 turn out (CCW).
Export to BADM