| ALL | | Anemometer || Barometer || Battery Bank || Camera || CH4 Probe || CO2 Probe || CO2 Transmitter || Conductivity Sensor || Data Logger || DO Sensor || Fluorometer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || LED sensor || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Water Level |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
PAR Sensor
2018-10-25 010159 PAR-LITE Installed PAR Sensor Incoming Installed at MB in the incoming position to replace sn 10162, which was brought back to the lab for repair/calibration.
2011-03-01 010161 PAR-LITE Installed PAR Sensor Reflected Water Installed on the Mayberry tule tower to measure PAR relfected from open water.
2018-10-25 010161 PAR-LITE Removed PAR Sensor Reflected Water Removed from MB (measuring reflected water) and brought back to the lab to replace the rotted cable and for calibration. It was replaced by sn 040412.
2019-07-25 010161 PAR-LITE Installed PAR Sensor Reflected Water Installed at MB (PAR reflected water) to replace sn 040412, which was brought back to the lab for testing.
2024-01-11 010161 PAR-LITE Removed PAR Sensor Reflected Water Removed from MB tower because it was no longer measuring open water (wetland is filling in with cattails)
2011-03-01 010163 PAR-LITE Installed PAR Sensor Reflected Veg Installed on the Mayberry tule tower measuring PAR reflected from tules.
2015-01-15 010163 PAR-LITE Removed PAR Sensor Reflected Veg Removed from the Mayberry tower (tule position) for repair due to INF reading.
2015-03-05 010163 PAR-LITE Installed PAR Sensor Reflected Veg Installed on the Mayberry tower (over tules) but it was reading high when it was connected to the multiplexor. The mux channel was even read high when jumped - so may be bad channel on mux. The PAR sensor was left on the tower and the mux or program will be changed to deal with the bad channel.
2018-10-25 040412 PAR-LITE Installed PAR Sensor Reflected Water Installed at MB (measuring reflected water) to replace sn 010161, which was brought back to the lab for repair/calibration.
2019-07-25 040412 PAR-LITE Removed PAR Sensor Reflected Water Removed from MB (PAR reflected water) and brought back to the lab for testing (spiky). It was replaced by sn 010161.
2011-03-01 10162 PAR-LITE Installed PAR Sensor Incoming Installed on the Mayberry tule tower measuring incoming PAR.
2018-10-25 10162 PAR-LITE Removed PAR Sensor Incoming Removed from MB and brought back to the lab to replace the rotted cable and for calibration. It was replaced by sn 010159.
2021-10-21 Q100420 LI-190/R Installed PAR Sensor Diffuse PAR Installed as part of the rotating shadow band for diffuse PAR - not working until 11-04
2024-08-15 UWQ11698 LI-192 Installed none Installed on the eastern edge of some cattails bout 5m east of the tower and about 50cm below the current water level. It is wired to an extension box about 3m from the tower.