Equipment and Calibration

Equipment history at Mayberry

| ALL | | Anemometer || Barometer || Battery Bank || Camera || CH4 Probe || CO2 Probe || CO2 Transmitter || Conductivity Sensor || Data Logger || DO Sensor || Fluorometer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || LED sensor || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Water Level |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
Multiplexer
2018-01-11 1236 AM16/32A Installed Multiplexer Met Installed on the Mayberry CR1000 met data logger to replace the existing AM16/32. There have been spikes in the data from the sensors on the existing multiplexer.
2018-02-28 1236 AM16/32A Removed Multiplexer Met Removed from Mayberry due to spikes and drops on radiation channels. This is an old mux and may be wearing out.
2017-01-26 1430 AM16/32A Installed Multiplexer Met Installed in new enclosure when switching from CR23X and CR1000.
2018-01-11 1430 AM16/32A Removed Multiplexer Met Removed from the Mayberry Tower where is was on the CR1000 met data logger. It was removed due to spikes in mostly the radiometer channels.
2011-03-01 2424 AM25T Installed Multiplexer Thermocouple Installed at the new Mayberry Tule Tower in about 1m of mater near a tule patch about 65m north east of the Mayberry tower.
2017-01-26 2424 AM25T Rewired Multiplexer Thermocouple Moved and re-wired AM25T from an old enclosure to a new enclosure when switching from CR23X and CR1000.
2018-02-28 26950 AM16/32B Installed Multiplexer Met This is a brand new mux. It was Installed on the Mayberry tower to replace the old one that seems to be causing spikes and drop out on the radiation channels.
2011-03-01 3052 AM16/32B Installed Multiplexer Met Installed at the new Mayberry Tule Tower in about 1m of mater near a tule patch about 65m north east of the Mayberry tower.
2017-01-26 3052 AM16/32B Removed Multiplexer Met Uninstalled when switching from CR23X and CR1000.
Export to BADM