Equipment and Calibration
Equipment history at Mayberry| ALL | | Anemometer || Barometer || Battery Bank || Camera || CH4 Probe || CO2 Probe || CO2 Transmitter || Conductivity Sensor || Data Logger || DO Sensor || Fluorometer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || LED sensor || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Water Level |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
Interface | |||||
2011-04-18 | 000196 | 1189PC | Removed | Removed from the Mayberry tower where it was sitting idle after being replaced with a RS422 to RS232 converter. | |
2016-04-13 | 3270 | A547 | Removed | Interface A547 | Removed from Mayberry with conductivity probe sn 6097 due to bad readings. Needs to be tested in the lab. |
2020-12-09 | 547-01 | A547 | Removed | Interface A547 | Removed from MB along with conductivity sensor sn 5688. Temperature reading is high so brought back to the lab for testing. |
2020-12-09 | 547-02 | A547 | Installed | Interface A547 | Installed with conductivity sensor sn 7157 at MB. |
2012-04-24 | 6763 | NL115 | Removed | Interface NL115 | Removed from the Mayberry eddy tower. No longer needed |
2013-02-14 | AIU-0397 | LI-7550 | Installed | Gas Analyzer CO2/H2O | Installed to replace sn 0614 for upgrade at Licor |
2013-11-21 | AIU-0397 | LI-7550 | Upgrade | Gas Analyzer CO2/H2O | The firmware of the 7550 box was upgraded from v6.5.2 to v6.5.5 to fix bugs with the USB memory overwrite and synchronizing the 7700 clock. |
2014-03-07 | AIU-0397 | LI-7550 | Upgrade | Gas Analyzer CO2/H2O | Updated the 7550 firmware from v6.5.5 to v7.0.1. Improved interface, support for SmartFlux, bug fixes. |
2014-04-28 | AIU-0397 | LI-7550 | Removed | Gas Analyzer CO2/H2O | Removed from the permanent Mayberry tower with the wash box and installed on the portable tower. Replaced by Ameriflux sn AIU-0915 |
2014-04-28 | AIU-0397 | LI-7550 | Installed | Gas Analyzer CO2/H2O | Installed on the Mayberry Portable tower with LI-7500 sn 75H-2176 and LI-7700 sn TG1-0224. This tower is mounted in the back of Dennis's truck and will be moved between two locations on a roughly weekly basis. |
2014-12-23 | AIU-0614 | LI-7550 | Installed | Gas Analyzer CO2/H2O | Installed on the Mayberry tower to replace sn AIU-1285 for painting. |
2015-01-15 | AIU-0614 | LI-7550 | Removed | Gas Analyzer CO2/H2O | Removed from tower and replaced by sn AIU-1285. This box had the wrong head on it and was brought back in for thorough forensics. |
2021-12-01 | AIU-0614 | LI-7550 | Installed | Interface LI-7550 | Installed at Mayberry to replace sn AIU-1285, which was removed for testing in the lab. |
2012-04-24 | AIU-0666 | LI-7550 | Installed | Gas Analyzer CO2/H2O | Installed on the Mayberry eddy tower to swap out LI-7500 sn 1594 |
2012-10-30 | AIU-0666 | LI-7550 | Removed | Gas Analyzer CO2/H2O | Removed from the Mayberry Tower due to corrosion on the connector circuit board. Replaced by sn 0614 |
2014-04-28 | AIU-0915 | LI-7550 | Installed | Interface LI-7550 | This is a borrowed sensor. It was installed on the Mayberry permanent tower to replace sn AIU-0397 for use on the portable tower. |
2014-08-19 | AIU-0915 | LI-7550 | Removed | Interface LI-7550 | This is part of the emergency response GHG system borrowed from Ameriflux. Removed from the Mayberry Tower with 7700 sn TG1-0352 and 7500 sn 75H-2339 and returned to Stephen at LBNL Ameriflux. |
2014-08-19 | AIU-1285 | LI-7550 | Installed | Interface LI-7550 | Installed on the Mayberry tower with 7700 sn TG1-0224 and 7500 sn 75H-2514 to replace the Ameriflux emergency response GHG system. |
2014-10-02 | AIU-1285 | LI-7550 | Upgrade | Interface LI-7550 | Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes. |
2014-12-23 | AIU-1285 | LI-7550 | Removed | Interface LI-7550 | Removed from the Mayberry tower for painting. Replaced by sn AIU-0614. |
2015-01-15 | AIU-1285 | LI-7550 | Installed | Interface LI-7550 | Installed on the Mayberry tower to replace sn AIU-1285. The existing box had a mismatched head. |
2016-04-13 | AIU-1285 | LI-7550 | Upgrade | Interface LI-7550 | Updated the 7550 firmware from v7.3.0 to v8.0.0 skiping version 7.5.0 Relevant 7.5.0 changes: Main Window is now resizable, Added serial number entry for sonics, Help update, Fixed timeout when saving config files Relevant 8.0.0 changes: Separate bandwidth settings for analog inputs and gas channels, Detection of new RS instruments, Additional calibration coefficients, Added sonic software version entry, Added chopper temp logging |
2017-08-16 | AIU-1285 | LI-7550 | Upgrade | Interface LI-7550 | The firmware was upgraded from v8.0.0 to v8.7.5 |
2017-11-30 | AIU-1285 | LI-7550 | Upgrade | Interface LI-7550 | Downgraded from embedded software version 8.7.5 to version 8.5.0. This solves the problem of data-header mismatch we had with version 8.7.5. |
2021-05-26 | AIU-1285 | LI-7550 | Upgrade | Interface LI-7550 | Upgraded firmware from v8.5.0 to v8.8.32 |
2021-12-01 | AIU-1285 | LI-7550 | Removed | Interface LI-7550 | Removed from Mayberry for testing and brought back to the lab. There were persistent error messages that prevented me from interacting with the software. Replaced by sn AIU-0614. |