Equipment history at Mayberry| ALL | | Anemometer || Barometer || Battery Bank || Camera || CH4 Probe || CO2 Probe || CO2 Transmitter || Conductivity Sensor || Data Logger || DO Sensor || Fluorometer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || LED sensor || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Water Level |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
Fluorometer | |||||
2017-03-29 | 2300565 | 2300-000 | Installed | Fluorometer | Installed at MB on the water sensor boom. It is at ?? depth below the water surface, about level with the top miniDOT DO and eosense CO2 sensors. |
2019-02-22 | 2300565 | 2300-000 | Removed | Fluorometer | Removed from MB and brought back to the lab to swap out its brushes and other maintenace. |
2019-02-28 | 2300565 | 2300-000 | Installed | Fluorometer | Reinstalled after a thorough cleaning and getting new brushes in the lab. |
2019-05-31 | 2300565 | 2300-000 | Removed | Fluorometer | Removed from the tower for a thorough cleaning in the lab. |
2019-06-04 | 2300565 | 2300-000 | Installed | Fluorometer | Installed at MB after a thorough cleaning in the lab. |
2020-02-20 | 2300565 | 2300-000 | Removed | Fluorometer | Removed from MB and brought back to the lab because its CDOM doesn't seem to be responding and it needs a good cleaning. |
2020-06-16 | 2300565 | 2300-000 | Installed | Fluorometer | Installed on MB floating boom. The power wires are backwards-- Black is +12V and white is G. |
2020-08-28 | 2300565 | 2300-000 | Removed | Fluorometer | Removed from MB floating boom and brought back to lab for a thorough cleaning |
2020-09-10 | 2300565 | 2300-000 | Installed | Fluorometer | Reinstalled after thorough cleaning in the lab. |
2020-10-22 | 2300565 | 2300-000 | Maintenance | Fluorometer | Replaced all 3 brushes. One brush had splayed needles that blocked one of the lenses (the lens with the visible red light) |
2021-01-17 | 2300565 | 2300-000 | Removed | Fluorometer | Removed from MB and brought back to the lab for further testing b/c it has noisy data |