Reports
Contents
Title: | PAR comparison 09-2018 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Date: | 2018-09-28 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Data File: | PARcompare_20180928.csv PARcompare_20180928_10min.csv PARcompare_20180928_1min.csv |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Refers to: | 161073,140455,040412,010159 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
We set up 3 PAR sensors from the lab (test sensors) next to each other for a comparison with an Ameriflux-calibrated (reference) PAR sensor from 2018-09-28 to 2018-10-03. The 4 sensors were set up about 1m outside of a south-facing lab window on the 2nd floor. There are some instances of hysteresis, but in general the 4 sensors are pretty well matched. They were connected to a CR6, which recorded data every 10 seconds. The new calibration coefficients are below. The original experiment ran for 5 days. Using only the first 2 days of data yielded calibration coefficients that were less than 0.2% (0.002) different.
 Figure 1. Time series of PAR data. Figure 4. Example of hysteresis between the 4 PAR sensors.
Figure 2. Scatter plot of test sensors plotted against the Ameriflux sensor. All R2 values are >0.99. Only values >0.2 umol/m2/s are included to prevent to much weight from night-time data.
Figure 3. Scatter plot of test sensor mV readings against the Ameriflux sensor umol/m2/s readings. Again, only values >0.2 umol/m2*s are included to prevent to much weight from night-time data. I used the slope of the linear regression lines to calculate the new calibration coefficients.  Averaging the data from 10s to 1-min or 10-min averages did not alter the calibration factors very much, so in the future 1-min or 10-min data could be used to calculate calibration factors if 10s data was not available.
Figure 6. Time series of PAR data averaged over 1 minute. Some hysteresis remains.
Figure 7. Scatter plot of 1-minute data plotted against the Ameriflux sensor umol/m2/s readings. Figure 5. Time series of PAR data averaged over 10 minutes. Hysteresis is gone. Figure 8. Scatter plot of 10-minute data plotted against the Ameriflux sensor umol/m2/s readings. |
|