Tonzi: 2004-08-27 10:45:00 (DOY 240)
Author: Siyan Ma
Others: Lorraine

Summary: 2 days lost on Floor eddy, Tower co2 profile cell temp low, no moisture point readings - dead battery,

When: Date: Aug. 27, 2004, Day: 240, Watch Time: 10:45 am
Where: Tonzi/Vaira Ranch
Who: Siyan and Lorraine
Weather: Sunny, hot, breeze on the tower
Phenology: dead grass; Oak leaves more yellow.

Summary & Problems

  1. Tower site computer was frozen on D238, so we lost 2-day data on Tower site. Floor and Vaira computer worked well and no data gap.
  2. Solar panel’s Amps = 0.0 amps. Something wrong with the power system?
  3. CO2 profile cell temp around 35 oC (< 50oC).
  4. TDR instrument didn’t work. The readings were always the same in the range 31 – 39 for different probes in Vaira and Tonzi. Voltage = 11.1v. Is Battery dead?
  5. Floor Laptop screen was flashing frequently.

Tonzi_Tower Site

  1. Fence and gate power: >12.0 v.
  2. Panel and Batteries: 14.2 v; 0.0 amps; 4.0 amps.
  3. When we got the site, the PC laptop wasn’t on. Frozen file: TW4_238.flx
  4. Reading from Gillsonic screen after restart. (The value of H2O and CO2 should be in the range of 0 – 5 v.)
    Time (Restart) 11:10:35
    Path/File name TW_240.flx
    mean std mean std
    w -0.65 0.848 Ts 34.848 2.135
    u 1.84 0.882 H2O 1.512 0.035
    v 3.53 1.077 CO2 1.5 0.018
  1. Download data. (Alt_X to close Gillsonic.exe; close PC208W before copying.)
  2. Check calibration cylinders:
    CO2 2000 psig; flow 8 psi.
    N2 900 psig; flow 10.0 psi.
  3. Checking CR10X and CR23X:
    CR10X OK. CR23X – NR_2 = -9999.

Toniz_Floor Site

  1. Fence and gate power: >12.0 v.
  2. Panel and Batteries: 13.7 v; 14.4 amps; 2.7 amps.
  3. Laptop screen was flashing frequently.
  4. Reading from Gillsonic screen before downloading data. (The value of H2O and CO2 should be in the range of 0 – 5 v.)
    Time 11:30:17
    Path/File name FL4_240.flx
    mean std mean std
    w 0.07 Ts 34.48
    u 0.71 H2O 1.87
    v 1.26 CO2 1.47
  1. Download data OK. Readings were OK from Gillsonic screen after downloading OK. The screen of PC was flashing.
  2. Check MD-9 system connections. OK.
  3. CO2 profile: cell temp still was lower than supposed 50oC.
  4. Close gate and set up fence.

Field Data

No tank data

No TZ Well data found

No TZ Moisture Point data found

No VR Moisture Point data found

No TZ grass heigth found

No VR grass heigth found

No TZ Tree DBH found

No water potential data found

No TZ Grass Biomass data found

No TZ OakLeaves Biomass data found

No TZ PineNeedles Biomass data found

0 photos found

Graphs display:
16 sets found

No data found for TZ_Fenergy.
11 TZ_tmet graphs found

Explore the graph:Tonzi Tower Met: Temperature and Power in a new window

Explore the graph:Tonzi Tower Met: Solar Radiation in a new window

Explore the graph:Tonzi Tower Met: IRT Sensors in a new window

Explore the graph:LED Sensor Incoming in a new window

Explore the graph:LED Sensor Outgoing in a new window

Explore the graph:4-Band Solar Radiation in a new window

Explore the graph:Air Temperature Profile in a new window

Explore the graph:Relativity Humidity Profile in a new window

Explore the graph:Tonzi Tower PRI in a new window

Explore the graph:SRS PRI/NDVI in a new window

Explore the graph:SRS NDVI in a new window

No data found for TZ_soilco2.
Can't check TZ_mixed
3 TZ_mixed graphs found

Explore the graph:Tonzi PPT in a new window

Explore the graph:Tonzi CO2 and H2O Densities in a new window

Explore the graph:Air Pressure in a new window

No data found for TZ_fflux.
8 TZ_fmet graphs found

Explore the graph:Soil Heat Flux in a new window

Explore the graph:Tonzi Soil Temperature Set A in a new window

Explore the graph:Tonzi Soil Temperatures Set B in a new window

Explore the graph:Tonzi Soil Temperatures Set C in a new window

Explore the graph:Tonzi Soil Temperature Set D in a new window

Explore the graph:Well Level in a new window

Explore the graph:Well Water Temperature in a new window

Explore the graph:Soil Water Potential in a new window

No data found for TZ_tram.
No data found for TZ_co2prof.
No data found for TZ_irt.
No data found for TZ_windprof.
No data found for TZ_tflux.
2 TZ_sm graphs found

Explore the graph:Floor Solar Radiations in a new window

Explore the graph:TZ Soil Moisture in a new window

Can't check TZ_met
1 TZ_met graphs found

Explore the graph:Floor and Tower RH in a new window

Can't check TZVR_mixed
No data found for TZ_osu.
No data found for TZ_fd.
Tonzi Z-cam data

plot of Tonzi z-cam LAI

plot of Tonzi z-cam gap fration