Equipment and Calibration History

View Equipment |

View history entry 965 for Licor LI-6262 sn: IRG3-1079

Date: 2013-04-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibrated in differential mode for check of new secondary standard. Ambient temperature was 28.8C, pressure was 100.55kPa. Using N2 (1100psi) and god bottle 402.69ppm CO2 (1800psi). Flowing N2 in both sample and reference CO2 +3ppm - adjusted to 0.0 H2O -0.19 - railed Flowing N2 in reference and god bottle in sample CO2 = 401.5ppm adjusted to 402.69ppm. Flowing N2 in reference and new secondary standard in sample, I recorded serial data with Hyperterm and took an average of 587 points. The new tank is 404.02ppm
Entered by: