Equipment and Calibration History

View Equipment |

View history entry 748 for Los Gatos Research FGGA sn: 09-0011

Date: 2012-07-20
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2011-10-06 Cell Pressure=138.44Torr, Cell Temperature=21.50C, Ring Time=13.448usec CH4 in air at 500psi CH4 Span=3.228ppm, CO2 Span=410ppm CH4 = 3.228 -> 3.228 Offset: 0ppb CO2 = 413 -> 410 Offset: -3ppm Notes: CH4 was right on and not calibrated. Only calibrated for CO2.
Entered by: