View history entry 672 for Licor LI-7500 sn: 0976

Date: 2011-01-03
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 0000-00-00 AGC=57 SDM=0 DAC1=0-1000 DAC2=10-25 Ambient P=100.6kPa, Tair=20.3C, Tdew=0C zero air at 0psi, CO2 597.4 at 0 Tdewspan=11.2 CO2 zero 0 -> 0.8594 Offset: 5.6ppm CO2 span 0 -> 1.0077 Offset: 3.9ppm H2O zero 0 -> 0.6298 Offset: -0.53ppm H2O span 0 -> 1.0139 Offset: 2.04ppm Notes: performed by Siyan
Entered by: