View history entry 6667 for YSI EXO fDOM sn: 22E100649

Date: 2024-04-08
From: Lab - none
To: Lab - none
Title: Calibration
Notes: fDOM (QSU and RFU), Last cal: 2024-01-23 EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-0.54 QSU,0.00 QSU,0.00 QSU,22.695 °C Pt2,300.00 QSU,132.47 QSU,300.02 QSU,0.00 QSU,22.144 °C Notes: Fresh calibration solution but pre-cal value is still way off. Maybe because of a dirty/cloudy sensor. This sensor probably needs to go back to the factory. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,-.31 RFU,0.00 RFU,0.00 RFU,22.796 °C Pt2,100.0 RFU,RFU,41.74 RFU,100.00 RFU,0.00 RFU,22.184 °C Notes: Fresh calibration solution but pre-cal value is still way off. Maybe because of a dirty/cloudy sensor. This sensor probably needs to go back to the factory.