Equipment and Calibration History

View Equipment |

View history entry 6646 for Licor LI-7500A sn: 75H-2669

Date: 2024-04-10
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2023-12-20 AGC=99.6 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.6kPa, Tair=26.04C, Tdew=12.04C Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00 CO2 zero 0.9187 -> .9189 Offset: .59ppm CO2 span 1.0167 -> 1.0177 Offset: -.65ppm H2O zero 0.9027 -> .9012 Offset: -8.8ppm H2O span 1.0206 -> 1.0115 Offset: .2ppm Notes:
Entered by: