View history entry 6645 for Licor LI-7500A sn: 75H-2667

Date: 2024-04-10
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2023-12-20 AGC=99.6 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.69kPa, Tair=24.8C, Tdew=11.0C Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00 CO2 zero 0.9413 -> .9409 Offset: -1.55ppm CO2 span 1.0207 -> 1.0213 Offset: -.35ppm H2O zero 0.8427 -> .8441 Offset: 8.5ppm H2O span 1.0363 -> 1.0404 Offset: -.09ppm Notes: Used 524.7 as the CO2 span point
Entered by: