View history entry 6625 for Licor LI-7500A sn: 75H-2180

Date: 2024-03-23
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-01-03 AGC=101.4 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.0kPa, Tair=24.65C, Tdew=11.42C Zero Air at 300psi, AMP CA04119 524.77 at 550 Tdewspan=20.00 CO2 zero 0.9141 -> .9137 Offset: -1.89ppm CO2 span 1.0406 -> 1.0356 Offset: -2.97ppm H2O zero 0.8965 -> .8951 Offset: -8.2ppm H2O span 1.0383 -> 1.0393 Offset: -.01ppm Notes: Used "524.52" as span point in software, otherwise, the post-span reading was too high