View history entry 6496 for Licor LI-7500A sn: 0035

Date: 2024-01-02
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2023-08-04 AGC=100.5 SDM=0 DAC1=- DAC2=- AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=101.0kPa, Tair=19.96C, Tdew=8.97C Zero Air at 550psi, AMP CA04119 524.77 at 750 Tdewspan=18.0 CO2 zero 0.9377 -> 0.9372 Offset: -2.0ppm CO2 span 1.024 -> 1.0267 Offset: -1.77ppm H2O zero 0.8984 -> 0.9006 Offset: 13.4ppm H2O span 1.0205 -> 1.0400 Offset: -0.4ppm Notes: I painted some scratches on the body with nail polish