Equipment and Calibration History

View Equipment |

View history entry 6491 for Licor LI-7500A sn: 75H-2669

Date: 2023-12-20
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2023-08-04 AGC=100.0 SDM=0 DAC1=- DAC2=- AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=99.6kPa, Tair=22.4C, Tdew=15.48C Zero Air at 550psi, AMP CA04119 524.77 at 750 Tdewspan=18.0 CO2 zero 0.9195 -> 0.9187 Offset: -3.0ppm CO2 span 1.0167 -> 1.0167 Offset: -0.02ppm H2O zero 0.9009 -> 0.9027 Offset: 10.9ppm H2O span 1.0163 -> 1.0206 Offset: -0.09ppm Notes: Reset SS from 102.6
Entered by: