Equipment and Calibration History

View Equipment |

View history entry 6489 for Licor LI-7500A sn: 75H-2514

Date: 2023-12-20
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2023-07-10 AGC=100.0 SDM=0 DAC1=- DAC2=- AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=99.6kPa, Tair=20.4C, Tdew=11.0C Zero Air at 600psi, AMP CA04119 524.77 at 750 Tdewspan=18.0 CO2 zero 0.9049 -> 0.9095 Offset: -3.4ppm CO2 span 1.0199 -> 1.0207 Offset: -0.37ppm H2O zero 0.8426 -> 0.8461 Offset: 3.5ppm H2O span 1.0244 -> 0.9962 Offset: -0.03ppm Notes: Calibration after factory repair and calibration. Reset SS from 103.4. Added SD coefficients to configuration.
Entered by: