Equipment and Calibration History

View Equipment |

View history entry 5952 for Licor LI-7500A sn: 75H-2669

Date: 2022-08-23
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-12-09 AGC=95.8 SDM= DAC1= DAC2= Ambient P=100kPa, Tair=26.03C, Tdew=16.96C Zero Air at 300psi, AMP CA04119 524.77 at 1400 Tdewspan=20.00 CO2 zero 0.9186 -> .9188 Offset: .94ppm CO2 span 1.0133 -> 1.0083 Offset: 3.8ppm H2O zero 0.9008 -> .9040 Offset: 20.6ppm H2O span 1.0123 -> 1.0174 Offset: -.11ppm Notes: Calibration using AMP loaner 7550 sn AIU-1039
Entered by: