Equipment and Calibration History

View Equipment |

View history entry 567 for Licor LI-7500A sn: 0035

Date: 2012-05-16
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2011-09-14 AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.19kPa, Tair=23.04C, Tdew=8.4C dry air at 16psi, CO2 436.53 at 1500 Tdewspan=13.7 CO2 zero 0.9338 -> 0.9419 Offset: ppm CO2 span 1.0108 -> 0.8842 Offset: ppm H2O zero 0.7353 -> 1.0055 Offset: ppm H2O span 0.9843 -> 0.9990 Offset: ppm Notes: recently fixed by Li-cor; double check in the lab conditions; no calibration applied yet.