Equipment and Calibration History

View Equipment |

View history entry 495 for Druck PDCR 1830-8388 sn: 2421703

Date: 2012-01-04
From: Lab - none
To: Lab - none
Title: Tested
Notes: I tested the pressure sensor in the lab by lowering in ito a vertical column of ABS pipe filled with water. The sensor was lowered into the water at depths marked on the sensor's cable, then removed from the water at the same depths: 0, 25, 50, 75, 100, 125, 100, 75, 50, 25, and 0cm The first test gave slightly higher readings than the second, but the third was between the first two, so I think all's well. depth in meters = 1.4504 * mV - 0.1881 R2 = 0.9952
Entered by: