Equipment and Calibration History

View Equipment |

View history entry 4394 for Licor LI-7500A sn: 0035

Date: 2020-02-19
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2019-11-18 AGC=100 SDM= DAC1= DAC2= Ambient P=100.95kPa, Tair=25.6C, Tdew=8.3C Zero Air at 800psi, AMP CA4396 408.97 at 550 Tdewspan=16.00 CO2 zero 0.9374 -> .9373 Offset: -.29ppm CO2 span 1.017 -> 1.0158 Offset: .43ppm H2O zero 0.898 -> .8981 Offset: .4ppm H2O span 1.0314 -> 1.0246 Offset: .14ppm Notes:
Entered by: