View history entry 4368 for Licor LI-7500A sn: 1594

Date: 2020-02-11
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2019-10-30 AGC=101.4 SDM=0 DAC1=- DAC2=- Ambient P=100.8kPa, Tair=26.1C, Tdew=-4.92C Zero Air at 800psi, AMP CA04396 408.97 at 550 Tdewspan=16.0 CO2 zero 0.8646 -> 0.8657 Offset: 4.6ppm CO2 span 1.0166 -> 1.0203 Offset: -1.92ppm H2O zero 0.8895 -> 0.8831 Offset: -18.3ppm H2O span 0.9958 -> 1.0120 Offset: -0.34ppm Notes: Ambient air very dry. Zero offsets a bit larger than normal.