View history entry 4359 for Licor LI-7500A sn: 75H-2669

Date: 2020-02-04
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2019-08-01 AGC=102.0 SDM=0 DAC1=- DAC2=- Ambient P=101.3kPa, Tair=25.6C, Tdew=-5.2C Zero Air at 850psi, AMP CA04396 408.97 at 550 Tdewspan=16.0 CO2 zero 0.9178 -> 0.9184 Offset: 2.3ppm CO2 span 1.0125 -> 1.0160 Offset: -1.87ppm H2O zero 0.9019 -> 0.9006 Offset: -8.4ppm H2O span 1.015 -> 1.0101 Offset: 0.1ppm Notes:
Entered by: