Equipment and Calibration History

View Equipment |

View history entry 3941 for Licor LI-7500A sn: 0035

Date: 2019-04-26
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-01-28 AGC=99.2 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=22.6C, Tdew=10.7C Zero Air at 1100psi, AMP CA4396 408.97 at 700 Tdewspan=16.00 CO2 zero 0.9375 -> .9371 Offset: -1.79ppm CO2 span 1.0193 -> 1.0189 Offset: .25ppm H2O zero 0.8969 -> .8989 Offset: 12.3ppm H2O span 1.0023 -> 1.0186 Offset: -.34ppm Notes:
Entered by: