Date: 2019-04-11
From: West Pond - Gas Analyzer CH4
To: West Pond - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: West Pond Last calibration: 2018-10-25 RSSI=71%, Optical RH=24% Ambient P=101.8kPa, Tair=17.2C Zero Air M4002 at 1600psi, AMP CA03923 3.1210 at 1700 CH4 Zero Offset: -0.005ppm CH4 Span Offset: 0.008ppm Notes: Windy day. Optical RH high - needs new desiccant.
Entered by:
Field notes