Equipment and Calibration History

View Equipment |

View history entry 3859 for Licor LI-7500A sn: 1594

Date: 2019-01-30
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-10-03 AGC=98.88 SDM=0 DAC1=0 DAC2=0 Ambient P=100.07kPa, Tair=27.07C, Tdew=10.85C Zero Air at 1200psi, AMP Air 408.97 at 825 Tdewspan=16.00 CO2 zero 0.8644 -> .8643 Offset: -.16ppm CO2 span 1.015 -> 1.0157 Offset: .5ppm H2O zero 0.887 -> .8891 Offset: 11.9ppm H2O span 1.0056 -> 1.0043 Offset: .02ppm Notes: Calibrated by Jiangong.
Entered by: