Equipment and Calibration History

View Equipment |

View history entry 3835 for Licor LI-7500A sn: 75H-2514

Date: 2019-01-07
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-09-26 AGC=102.5 SDM=0 DAC1=0 DAC2=0 Ambient P=100.9kPa, Tair=18.0C, Tdew=10.1C Zero Air at 1200psi, AMP Air 408.97 at 850 Tdewspan=16.00 CO2 zero 0.9045 -> .9051 Offset: 2.36ppm CO2 span 1.0036 -> 1.0167 Offset: -6.83ppm H2O zero 0.837 -> .8397 Offset: 19.5ppm H2O span 1.0165 -> 1.0177 Offset: -.03ppm Notes:
Entered by: