View history entry 356 for Los Gatos Research FGGA sn: 09-0011

Date: 2011-05-13
From: Lab - none
To: Lab - none
Title: Maintenance
Notes: On first power up CH4 and H2O did not line lock. I adjusted the VO setting in the linelock_0.ini file from -0.690 to -0.590. I did a calibration with the CH4 god bottle 1.8978ppm CH4, 350psi. Last cal was: CH4 - 14 Jul 2010 H2O, CO2 - 13 Mar 2009 Precal: 14:26PST, 22.77C, 138.11Torr, 14.052usec, 1Hz, 2.058ppm After cal: 1.9021ppm