Equipment and Calibration History

View Equipment |

View history entry 316 for Licor LI-7700 sn: TG1-0100

Date: 2010-05-05
From: Lab - none
To: Twitchell Rice - Gas Analyzer CH4
Title: Installed
Notes: Received from Liukang Xu of licor. Installed on the Twitchell tower. Single ended analog outs are being used to record the LI7700 data on the CR1000: Channel 0V 5V wire CH4 (mmol/m3) 0.05 0.15 wht CH4 (ppm) 1 5 brn Pressure (kPa) 90 110 grn Temperature (C) 0 50 yel RSSI (%) 0 100 gry The mirrors are set to clean at an RSSI value of 20%. USB logging is set up at 10Hz.
Entered by: