View history entry 3140 for Licor LI-7500A sn: 75H-2150

Date: 2017-10-11
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-05-30 AGC=96.8 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=23.9C, Tdew=5.5C Zero Air at 500psi, AMP CO2 408.97 at 2000 Tdewspan=16.0 CO2 zero 0.9024 -> 0.9018 Offset: -2.42ppm CO2 span 1.0194 -> 1.0070 Offset: 6.38ppm H2O zero 0.8664 -> 0.8649 Offset: -10.1ppm H2O span 0.9975 -> 1.0090 Offset: 0.24ppm Notes:
Entered by: