Equipment and Calibration History

View Equipment |

View history entry 3042 for Licor LI-7500A sn: 0035

Date: 2017-09-05
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-05-08 AGC=99.8 SDM=0 DAC1=0 DAC2=0 Ambient P=100.4kPa, Tair=29.1C, Tdew=18.4C Zero Air at 1000psi, AMP Air 408.97 at 2100 Tdewspan=16.00 CO2 zero 0.9381 -> .9373 Offset: -3.34ppm CO2 span 1.023 -> 1.0084 Offset: 7.67ppm H2O zero 0.8974 -> .8976 Offset: .8ppm H2O span 1.0261 -> 1.0141 Offset: .25ppm Notes:
Entered by: