Equipment and Calibration History

View Equipment |

View history entry 3038 for Licor LI-7500A sn: 75H-2182

Date: 2017-08-16
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-04-20 AGC=87.5 SDM=0 DAC1=0 DAC2=0 Ambient P=100.4kPa, Tair=23.9C, Tdew=12.7C Zero Air at 1075psi, AMP Air 408.97 at 2000 Tdewspan=16.00 CO2 zero 0.8838 -> .8832 Offset: -2.73ppm CO2 span 1.0164 -> 1.0123 Offset: 2.13ppm H2O zero 0.8454 -> .8436 Offset: -11.4ppm H2O span 1.0139 -> 1.0038 Offset: .2ppm Notes:
Entered by: