Date: 2010-06-16
From: Twitchell Rice - Gas Analyzer CH4
To: Twitchell Rice - Gas Analyzer CH4
Title: Calibration
Notes: Clibrated in the field with grade 5.0 ultra pure N2 and 3.276ppm CH4 secondary standard. zero before: ~0.009 ppm zero after: ~0.000 ppm span before: ~3.333 ppm span after: ~3.277 ppm
Entered by: