View history entry 2795 for Licor LI-7500A sn: 1594

Date: 2017-05-30
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-01-27 AGC=101.4 SDM= DAC1= DAC2= Ambient P=99.7kPa, Tair=19.3C, Tdew=9.5C Zero Air at 1050psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.8654 -> .8653 Offset: .79ppm CO2 span 1.0194 -> 1.0187 Offset: 0.42ppm H2O zero 0.8868 -> .8861 Offset: -4.1ppm H2O span 1.0067 -> 1.0256 Offset: -.39ppm Notes:
Entered by: