Equipment and Calibration History

View Equipment |

View history entry 2753 for Licor LI-7500A sn: 0035

Date: 2017-05-08
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-12-05 AGC=100 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=21.9C, Tdew=7.8C Zero Air at 1550psi, LBL Air 513.22 at 350 Tdewspan=16.00 CO2 zero 0.939 -> .9381 Offset: -3.83ppm CO2 span 1.0265 -> 1.0230 Offset: 2.41ppm H2O zero 0.9025 -> .8974 Offset: -28.2ppm H2O span 1.0029 -> 1.0261 Offset: -.48ppm Notes: First calibration since upgrade from 7500 to 7500A.
Entered by: