View history entry 2096 for Licor LI-7500A sn: 75H-2182

Date: 2015-11-10
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2015-07-15 AGC=87.4 SDM=6 DAC1=- DAC2=- Ambient P=100.94kPa, Tair=21.68C, Tdew=6.22C Zero Air at 700psi, God CO2 421.22 at 700 Tdewspan=11.08 CO2 zero 0.8825 -> 0.8833 Offset: 3.48ppm CO2 span 1.0064 -> 1.0197 Offset: -7.02ppm H2O zero 0.8458 -> 0.8464 Offset: 4.0ppm H2O span 1.0016 -> 1.0242 Offset: -0.43ppm Notes: