View history entry 1908 for Licor LI-7500A sn: 75H-2182

Date: 2015-07-15
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2015-01-13 AGC=84.3 SDM=1 DAC1=- DAC2=- Ambient P=100.19kPa, Tair=22.31C, Tdew=15.05C Zero Air at 800psi, God CO2 421.22 at 850 Tdewspan=11.08 CO2 zero 0.8824 -> 0.8825 Offset: 0.43ppm CO2 span 1.0118 -> 1.0064 Offset: 2.88ppm H2O zero 0.8453 -> 0.8458 Offset: 2.9ppm H2O span 1.0078 -> 1.0016 Offset: 0.12ppm Notes:
Entered by: