View history entry 1786 for Licor LI-7500A sn: 1594

Date: 2015-04-20
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2014-12-22 AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.28kPa, Tair=20.38C, Tdew=12.81C Zero Air at 900psi, God CO2 421.22 at 1200 Tdewspan=15.96 CO2 zero 0.8649 -> 0.8648 Offset: -0.44ppm CO2 span 1.0175 -> 1.0178 Offset: -0.22ppm H2O zero 0.8981 -> 0.8985 Offset: 2.2ppm H2O span 1.1398 -> 1.0087 Offset: 2.68ppm Notes: Relatively large offsets in H2O span, similar to other 7500s calibrated at this time - dew point generator and Vaisala probe agreed with in 0.5C