Equipment and Calibration History

View Equipment |

View history entry 1662 for Licor LI-7500A sn: 75H-2182

Date: 2015-01-13
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2014-08-25 AGC=85.3 SDM=- DAC1=- DAC2=- Ambient P=101.1kPa, Tair=25.52C, Tdew=8.45C Zero Air at 900psi, God CO2 421.22 at 1300 Tdewspan=15.95 CO2 zero 0.8816 -> 0.8824 Offset: 3.53ppm CO2 span 1.0074 -> 1.0118 Offset: -2.38ppm H2O zero 0.8463 -> 0.8453 Offset: -6.2ppm H2O span 1.0039 -> 1.0078 Offset: -0.08ppm Notes:
Entered by: