View history entry 1479 for Los Gatos Research FGGA sn: 09-0012

Date: 2010-02-18
From: Twitchell Rice -
To: Twitchell Rice -
Title: Calibration
Notes: Last calibration: 2010-01-14 00:00:00 Cell Pressure=39.54kPa, Cell Temperature=27.98C, Ring Time=17.17usec 2nd standard at 0psi: CH4: 3.276ppm, CO2: 433ppm, H2O: 0ppm CH4 = 3.5032ppm -> 3.316ppm Offset: -0.227ppb CO2 = 448.8ppm -> 441ppm Offset: -15.8ppm Notes: External filter seemed plugged up, swapped it with cleaned filter before calibration. Values improved, but I still don't agree with cal tank. I think the very low pressure is a problem. Maybe the internal filter is plugged up. I will do some check in the lab and then possibly swap out the LGR.
Entered by: Sara Knox