View history entry 1464 for Licor LI-7500A sn: 75H-2150

Date: 2014-06-24
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2013-12-18 AGC=95.08 SDM=1 DAC1=- DAC2=- Ambient P=100.15kPa, Tair=21.78C, Tdew=11.81C Zero Air at 1200psi, God CO2 421.22 at 1500 Tdewspan=15.26 CO2 zero 0.9012 -> 0.9012 Offset: 0.07ppm CO2 span 1.0016 -> 1.0011 Offset: 0.28ppm H2O zero 0.8643 -> 0.8652 Offset: 5.8ppm H2O span 0.9987 -> 1.0087 Offset: -0.21ppm Notes: