Equipment and Calibration History

View Equipment |

View history entry 1221 for Licor LI-7500A sn: 75H-2182

Date: 2014-02-28
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2013-08-20 AGC=83.8 SDM=1 DAC1=- DAC2=- Ambient P=98.10kPa, Tair=19.40C, Tdew=9.95C Zero air at 1400psi, God CO2 421.22 at 1500 Tdewspan=15.17 CO2 zero 0.8802 -> 0.8815 Offset: 5.98ppm CO2 span 0.9947 -> 1.0064 Offset: -4.84ppm H2O zero 0.8471 -> 0.8476 Offset: 2.5ppm H2O span 0.9854 -> 0.9875 Offset: -0.05ppm Notes: Using box sn 0666. Before calibration upgraded firmware from v6.5.5 to v7.0.1 - FPGA v6.0.3 not needing update.
Entered by: