Date: |
2013-05-20 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration: 2012-12-12
AGC=90 SDM=1 DAC1=- DAC2=-
Ambient P=100.46kPa, Tair=27.2C, Tdew=12C
Zero air at 1800psi, CO2 in air 421.22 at 1700 Tdewspan=15
CO2 zero 0.9145 -> 0.9133 Offset: -5.41ppm
CO2 span 1.009 -> 1.0091 Offset: 0.07ppm
H2O zero 0.8788 -> 0.8840 Offset: 33.6ppm
H2O span 0.9989 -> 0.9991 Offset: 0.1ppm
Notes: Upgraded firmware to version 6.5.2 (on box SN 0674) and now using new interface version 6.5.0. Re-entered calibration coefficients from last factory calibration (12/12/12) since we forgot to save the config file last time. |
Entered by: |
|
|
|