Instrument:Water Level
Manufacture:METER Group
Model No:HYDROS21
Serial No:H21G00003238
Tag No:
Acquisition Date:2024-10-07
Project:Dutch Slough
Location on 2025-04-24:Lab,
Notes: Replacement under warranty for H21G200002433. EC, temperature, water depth, 10 meter Pigtail Cable

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
7196 2025-04-24 Lab none Tested Lab none Tested on a CR1000 in the lab. Seems like it is working fine. I think the failure in the field was do to the extension wires getting chewed or other external wiring issue. The sensor seems fine for reinstall.
7186 2025-04-08 Gilbert Tract Water Level Boardwalk Removed Lab none Disconnected that the junction box and brought in. This sensor stop working a couple of weeks ago.
6915 2024-10-10 Lab none Installed Gilbert Tract Water Level Boardwalk Installed on the fence post just off of the boardwalk walkway. It is co-located with the Campbell conductivity sensor.